Commands by Subsystem 3

Table 67 :WAVeform Commands Summary (continued)

Command

Query

Options and Query Returns

 

 

 

n/a

:WAVeform:YINCrement?

<return_value> ::= y-increment

 

(see page 431)

value in the current preamble in

 

 

NR3 format

 

 

 

n/a

:WAVeform:YORigin?

<return_value> ::= y-origin in

 

(see page 432)

the current preamble in NR3

 

 

format

 

 

 

n/a

:WAVeform:YREFerence?

<return_value> ::= y-reference

 

(see page 433)

value in the current preamble in

 

 

NR1 format

 

 

 

Introduction to

The WAVeform subsystem is used to transfer data to a controller from the

:WAVeform

oscilloscope waveform memories. The queries in this subsystem will only

Commands

operate when the channel selected by :WAVeform:SOURce is on.

 

Waveform Data and Preamble

 

The waveform record is actually contained in two portions: the preamble

 

and waveform data. The waveform record must be read from the

 

oscilloscope by the controller using two separate commands,

 

:WAVeform:DATA (see page 410) and :WAVeform:PREamble (see page 417).

 

The waveform data is the actual data acquired for each point in the

 

specified source. The preamble contains the information for interpreting

 

the waveform data, which includes the number of points acquired, the

 

format of acquired data, and the type of acquired data. The preamble also

 

contains the X and Y increments, origins, and references for the acquired

 

data, so that word and byte data can be translated to time and voltage

 

values.

Data Acquisition Types

There are three types of waveform acquisitions that can be selected for analog channels with the :ACQuire:TYPE command (see page 132): NORMal, AVERage, PEAK, and HRESolution. When the data is acquired using the :DIGitize command (see page 96) or :RUN command (see

page 116), the data is placed in the channel buffer of the specified source.

Once you have acquired data with the :DIGitize command, the instrument is stopped. If the instrument is restarted (via GPIB or the front panel), or if any instrument setting is changed, the data acquired with the :DIGitize command may be overwritten.You should first acquire the data with the :DIGitize command, then immediately read the data with the :WAVeform:DATA? query (see page 410) before changing any instrument setup.