:CHANnel<n>:PROBe:SKEW

(see page 530)

Command Syntax :CHANnel<n>:PROBe:SKEW <skew value> <skew value> ::= skew time in NR3 format <skew value> ::= -100 ns to +100 ns <n> ::= {1 2 3 4}

The :CHANnel<n>:PROBe:SKEW command sets the channel- to- channel skew factor for the specified channel. Each analog channel can be adjusted

+or - 100 ns for a total of 200 ns difference between channels. You can use the oscilloscope's probe skew control to remove cable- delay errors between channels.

Query Syntax :CHANnel<n>:PROBe:SKEW?

The :CHANnel<n>:PROBe:SKEW? query returns the current probe skew setting for the selected channel.

Return Format <skew value><NL>

<skew value> ::= skew value in NR3 format

See Also "Introduction to :CHANnel<n> Commands" on page 143