3Commands by Subsystem

Table 32 Status Byte Register (STB) (continued)

Bit

Name

Description

When Set (1 = High = True), Indicates:

 

 

 

 

6

RQS

Request Service

When polled, that the device is requesting service.

 

 

 

 

 

MSS

Master Summary Status

When read (by *STB?), whether the device has a

 

 

 

reason for requesting service.

 

 

 

 

5

ESB

Event Status Bit

An enabled condition in the Standard Event Status

 

 

 

Register (ESR) has occurred.

 

 

 

 

4

MAV

Message Available

There are messages in the Output Queue.

 

 

 

 

3

---

---

(Not used, always 0.)

 

 

 

 

2

MSG

Message

An advisory has been displayed on the

 

 

 

oscilloscope.

 

 

 

 

1

USR

User Event

An enabled user event condition has occurred.

 

 

 

 

0

TRG

Trigger

A trigger has occurred.

 

 

 

 

N O T E

To read the instrument's status byte with RQS reported on bit 6, use the interface Serial

Poll.

 

See Also "Introduction to Common (*) Commands" on page 63

 

"*SRE (Service Request Enable)" on page 79

82

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 82
Image 82