Commands by Subsystem 3

:SBUS:UART:FRAMing

(see page 530)

Command Syntax :SBUS:UART:FRAMing <value>

<value> ::= {OFF <decimal> <nondecimal>}

<decimal> ::= 8-bit integer in decimal from 0-255 (0x00-0xff) <nondecimal> ::= #Hnn where n ::= {0,..,9 A,..,F} for hexadecimal <nondecimal> ::= #Bnn...n where n ::= {0 1} for binary

 

 

The :SBUS:UART:FRAMing command determines the byte value to use for

 

 

framing (end of packet) or to turn off framing for UART decode.

 

 

This command is only valid on 4 (analog) channel oscilloscope models when the

N O T E

 

 

UART/RS-232 triggering and serial decode option (Option 232) has been licensed.

 

 

 

 

Query Syntax

:SBUS:UART:FRAMing?

 

 

The :SBUS:UART:FRAMing? query returns the current UART decode base

 

 

setting.

Return Format

<value><NL>

 

 

<value> ::= {OFF <decimal>}

 

 

<decimal> ::= 8-bit integer in decimal from 0-255

Errors

"- 241, Hardware missing" on page 501

See Also

"Introduction to :SBUS Commands" on page 277

 

 

":TRIGger:UART Commands" on page 387

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

293