Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 293

:SBUS:UART:FRAMing

(see page 530)
Command Syntax :SBUS:UART:FRAMing <value>
<value> ::= {OFF | <decimal> | <nondecimal>}
<decimal> ::= 8-bit integer in decimal from 0-255 (0x00-0xff)
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
The :SBUS:UART:FRAMing command determines the byte value to use for
framing (end of packet) or to turn off framing for UART decode.
Query Syntax :SBUS:UART:FRAMing?
The :SBUS:UART:FRAMing? query returns the current UART decode base
setting.
Return Format <value><NL>
<value> ::= {OFF | <decimal>}
<decimal> ::= 8-bit integer in decimal from 0-255
Errors "- 241, Hardware missing" on page 501
See Also "Introduction to :SBUS Commands" on page 277
":TRIGger:UART Commands" on page 387
NOTE This command is only valid on 4 (analog) channel oscilloscope models when the
UART/RS-232 triggering and serial decode option (Option 232) has been licensed.