Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 169

:DISPlay:SOURce

(see page 530)
Command Syntax :DISPlay:SOURce <value>
<value> ::= {PMEMory0 | PMEMory1 | PMEMory2 | PMEMory3 | PMEMory4
| PMEMory5 | PMEMory6 | PMEMory7 | PMEMory8 | PMEMory9}
PMEMory0-9 ::= pixel memory 0 through 9
The :DISPlay:SOURce command specifies the default source and
destination for the :DISPlay:DATA command and query. PMEMory0- 9
correspond to the INTERN_0- 9 files found in the front panel Save/Recall
menu.
Query Syntax :DISPlay:SOURce?
The :DISPlay:SOURce? query returns the specified SOURce.
Return Format <value><NL>
<value> ::= {PMEM0 | PMEM1 | PMEM2 | PMEM3 | PMEM4 | PMEM5 | PMEM6
| PMEM7 | PMEM8 | PMEM9}
See Also "Introduction to :DISPlay Commands" on page 161
":DISPlay:DATA" on page 164