Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 341

:TRIGger[:EDGE] Commands

Table 59 :TRIGger[:EDGE] Commands Summary

Command Query Options and Query Returns
:TRIGger[:EDGE]:COUPl
ing {AC | DC | LF}
(see page 342)
:TRIGger[:EDGE]:COUPl
ing? (see page 342)
{AC | DC | LF}
:TRIGger[:EDGE]:LEVel
<level> [,<source>]
(see page 343)
:TRIGger[:EDGE]:LEVel
? [<source>] (see
page 343)
For internal triggers, <level>
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers, <level>
::= 2 volts with probe
attenuation at 1:1 in NR3 format.
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format
:TRIGger[:EDGE]:REJec
t {OFF | LF | HF}
(see page 344)
:TRIGger[:EDGE]:REJec
t? (see page 344)
{OFF | LF | HF}
:TRIGger[:EDGE]:SLOPe
<polarity> (see
page 345)
:TRIGger[:EDGE]:SLOPe
? (see page 345)
<polarity> ::= {POSitive |
NEGative | EITHer | ALTernate}
:TRIGger[:EDGE]:SOURc
e <source> (see
page 346)
:TRIGger[:EDGE]:SOURc
e? (see page 346)
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format