Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 147

:CHANnel<n>:DISPlay

(see page 530)
Command Syntax :CHANnel<n>:DISPlay <display value>
<display value> ::= {{1 | ON} | {0 | OFF}}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:DISPlay command turns the display of the specified
channel on or off.
Query Syntax :CHANnel<n>:DISPlay?
The :CHANnel<n>:DISPlay? query returns the current display setting for
the specified channel.
Return Format <display value><NL>
<display value> ::= {1 | 0}
See Also "Introduction to :CHANnel<n> Commands" on page 143
":VIEW" on page 122
":BLANk" on page 94
":STATus" on page 119