5Obsolete and Discontinued Commands

:TIMebase:DELay

(see page 530)

Command Syntax :TIMebase:DELay <delay_value>

<delay_value> ::= time in seconds from trigger to the delay reference point on the screen.

The valid range for delay settings depends on the time/division setting for the main time base.

The :TIMebase:DELay command sets the main time base delay. This delay is the time between the trigger event and the delay reference point on the screen. The delay reference point is set with the :TIMebase:REFerence command (see page 307).

N O T E

The :TIMebase:DELay command is obsolete and is provided for backward compatibility to

previous oscilloscopes. Use the :TIMebase:POSition command (see page 305) instead.

 

Query Syntax :TIMebase:DELay?

The :TIMebase:DELay query returns the current delay value.

Return Format <delay_value><NL>

<delay_value> ::= time from trigger to display reference in seconds in NR3 format.

Example Code ' TIMEBASE_DELAY - Sets the time base delay. This delay

'is the internal time between the trigger event and the

'onscreen delay reference point.

'Set time base delay to 0.0.

myScope.WriteString ":TIMEBASE:DELAY 0.0"

Example program from the start: "VISA COM Example in Visual Basic" on page 578