Commands Quick Reference 2

Table 18 General :TRIGger Commands Summary (continued)

Command

Query

Options and Query Returns

 

 

 

:TRIGger:MODE <mode>

:TRIGger:MODE? (see

<mode> ::= {EDGE GLITch

(see page 319)

page 319)

PATTern DURation TV}

 

 

<return_value> ::= {<mode>

 

 

<none>}

 

 

<none> ::= query returns "NONE"

 

 

if the :TIMebase:MODE is ROLL or

 

 

XY

 

 

 

:TRIGger:NREJect {{0

:TRIGger:NREJect?

{0 1}

OFF} {1 ON}}

(see page 320)

 

(see page 320)

 

 

 

 

 

:TRIGger:PATTern

:TRIGger:PATTern?

<value> ::= integer in NR1 format

<value>, <mask>

(see page 321)

or <string>

[,<edge

 

<mask> ::= integer in NR1 format

source>,<edge>] (see

 

or <string>

page 321)

 

<string> ::= "0xnn"; n ::=

 

 

{0,..,9 A,..,F} (# bits = #

 

 

channels)

 

 

<edge source> ::= {CHANnel<n>

 

 

EXTernal NONE}

 

 

<edge> ::= {POSitive NEGative}

 

 

<n> ::= 1-2 or 1-4 in NR1 format

 

 

 

:TRIGger:SWEep

:TRIGger:SWEep? (see

<sweep> ::= {AUTO NORMal}

<sweep> (see

page 323)

 

page 323)

 

 

 

 

 

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

45

Page 45
Image 45
Agilent Technologies 5000 Series manual TRIGgerMODE mode TRIGgerMODE? see Mode = Edge GLITch