Commands by Subsystem 3

Table 33 Root (:) Commands Summary (continued)

Command

Query

Options and Query Returns

 

 

 

:OVLenable <mask>

:OVLenable? (see

<mask> ::= 16-bit integer in NR1

(see page 111)

page 112)

format as shown:

 

 

Bit Weight Input

 

 

--- ------

----------

 

 

10

1024

Ext Trigger Fault

 

 

9

512

Channel 4 Fault

 

 

8

256

Channel 3 Fault

 

 

7

128

Channel 2 Fault

 

 

6

64

Channel 1 Fault

 

 

4

16

Ext Trigger OVL

 

 

3

8

Channel 4 OVL

 

 

2

4

Channel 3 OVL

 

 

1

2

Channel 2 OVL

 

 

0

1

Channel 1 OVL

 

 

 

n/a

:OVLRegister? (see

<value> ::= integer in NR1

 

page 113)

format. See OVLenable for <value>

 

 

 

:PRINt [<options>]

n/a

<options> ::= [<print

(see page 115)

 

option>][,..,<print option>]

 

 

<print option> ::= {COLor

 

 

GRAYscale PRINter0 BMP8bit

 

 

BMP PNG NOFactors FACTors}

 

 

<print option> can be repeated up

 

 

to 5 times.

 

 

 

 

 

:RUN (see page 116)

n/a

n/a

 

 

 

 

 

n/a

:SERial (see

<return value> ::= unquoted

 

page 117)

string containing serial number

 

 

 

 

 

:SINGle (see

n/a

n/a

 

 

page 118)

 

 

 

 

 

 

 

 

n/a

:STATus? <display>

{0 1}

 

 

(see page 119)

<display> ::= {CHANnel<n>

 

 

FUNCtion MATH}

 

 

<n> ::= 1-2 or 1-4 in NR1 format

 

 

 

 

 

:STOP (see page 120)

n/a

n/a

 

 

 

 

 

 

n/a

:TER? (see page 121)

{0 1}

 

 

 

 

:VIEW <source> (see

n/a

<source> ::= {CHANnel<n>

page 122)

 

PMEMory{0 1 2 3 4 5 6

 

 

7 8 9} FUNCtion MATH}

 

 

<n> ::= 1-2 or 1-4 in NR1 format

 

 

 

 

 

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

87

Page 87
Image 87