Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 371

:TRIGger:LIN:TRIGger

(see page 530)
Command Syntax :TRIGger:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID}
The :TRIGger:LIN:TRIGger command sets the LIN trigger on condition to be
Sync Break (SYNCbreak) or Frame Id (ID).
Query Syntax :TRIGger:LIN:TRIGger?
The :TRIGger:LIN:TRIGger? query returns the current LIN trigger value.
Return Format <condition><NL>
<condition> ::= {SYNC | ID}
Errors "- 241, Hardware missing" on page 501
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:MODE" on page 319
":TRIGger:LIN:SIGNal:DEFinition" on page 497
":TRIGger:LIN:SOURce" on page 368
NOTE The ID option is available when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.