Obsolete and Discontinued Commands 5

:MEASure:TMIN

(see page 530)

Command Syntax :MEASure:TMIN [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

 

 

The :MEASure:TMIN command installs a screen measurement and starts an

 

 

X- at- Min- Y measurement on the selected waveform. If the optional source

 

 

is specified, the current source is modified.

 

 

The :MEASure:TMIN command is obsolete and is provided for backward compatibility to

N O T E

 

 

previous oscilloscopes. Use the :MEASure:XMIN command (see page 256) instead.

 

 

 

 

Query Syntax

:MEASure:TMIN? [<source>]

 

 

The :MEASure:TMIN? query returns the horizontal axis value at which the

 

 

minimum vertical value occurs on the current source. If the optional

 

 

source is specified, the current source is modified. If all channels are off,

 

 

the query returns 9.9E+37.

Return Format

<value><NL>

 

 

<value> ::= time at minimum in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

 

":MEASure:TMAX" on page 482

 

 

":MEASure:XMAX" on page 255

 

 

":MEASure:XMIN" on page 256

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

483

Page 483