3Commands by Subsystem

:TRIGger[:EDGE]:REJect

Command Syntax

N O T E

(see page 530)

:TRIGger[:EDGE]:REJect <reject>

<reject> ::= {OFF LFReject HFReject}

The :TRIGger[:EDGE]:REJect command turns the low- frequency or

high- frequency reject filter on or off. You can turn on one of these filters at a time.

The high frequency reject filter adds a 50 kHz low- pass filter in the trigger path to remove high frequency components from the trigger waveform. Use the high frequency reject filter to remove high- frequency noise, such as AM or FM broadcast stations, from the trigger path.

The low frequency reject filter adds a 50 kHz high- pass filter in series with the trigger waveform to remove any unwanted low frequency components from a trigger waveform, such as power line frequencies, that can interfere with proper triggering.

The :TRIGger[:EDGE]:REJect and the :TRIGger[:EDGE]:COUPling selections are coupled. Changing the setting of the :TRIGger[:EDGE]:COUPling can change the COUPling setting.

Query Syntax

:TRIGger[:EDGE]:REJect?

 

The :TRIGger[:EDGE]:REJect? query returns the current status of the reject

 

filter.

Return Format

<reject><NL>

 

<reject> ::= {OFF LFR HFR}

See Also

"Introduction to :TRIGger Commands" on page 313

 

":TRIGger:HFReject" on page 317

 

":TRIGger[:EDGE]:COUPling" on page 342

344

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 344
Image 344
Agilent Technologies 5000 Series manual TRIGgerEDGEREJect, TRIGgerEDGECOUPling on