Commands by Subsystem 3

:TIMebase:WINDow:RANGe

(see page 530)

Command Syntax :TIMebase:WINDow:RANGe <range value>

<range value> ::= range value in seconds in NR3 format

The :TIMebase:WINDow:RANGe command sets the full- scale horizontal time in seconds for the delayed window. The range is 10 times the current delayed view window seconds per division setting. The main sweep range determines the range for this command. The maximum value is one half of the :TIMebase:RANGe value.

Query Syntax :TIMebase:WINDow:RANGe?

The :TIMebase:WINDow:RANGe? query returns the current window timebase range setting.

Return Format <value><NL>

<value> ::= range value in seconds

See Also "Introduction to :TIMebase Commands" on page 302

":TIMebase:RANGe" on page 306

":TIMebase:POSition" on page 305

":TIMebase:SCALe" on page 308

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

311

Page 311
Image 311
Agilent Technologies 5000 Series manual TIMebaseWINDowRANGe, Return Format valueNL