Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 311

:TIMebase:WINDow:RANGe

(see page 530)
Command Syntax :TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full-scale horizontal time
in seconds for the delayed window. The range is 10 times the current
delayed view window seconds per division setting. The main sweep range
determines the range for this command. The maximum value is one half of
the :TIMebase:RANGe value.
Query Syntax :TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window
timebase range setting.
Return Format <value><NL>
<value> ::= range value in seconds
See Also "Introduction to :TIMebase Commands" on page 302
":TIMebase:RANGe" on page 306
":TIMebase:POSition" on page 305
":TIMebase:SCALe" on page 308