Obsolete and Discontinued Commands 5

:MEASure:TSTOp

(see page 530)

Command Syntax :MEASure:TSTOp <value> [suffix]

<value> ::= time at the stop marker in seconds [suffix] ::= {s ms us ns ps}

 

 

The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the

 

 

specified time with respect to the trigger time.

 

 

The short form of this command, TSTO, does not follow the defined Long Form to Short

N O T E

 

 

Form Truncation Rules (see page 532). The normal short form "TST" would be the same for

 

 

 

 

both TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.

 

 

 

 

 

The :MEASure:TSTOp command is an obsolete command provided for compatibility to

N O T E

 

 

previous oscilloscopes. Use the :MARKer:X2Position command (see page 208) instead.

 

 

 

 

Query Syntax

:MEASure:TSTOp?

 

 

The :MEASure:TSTOp? query returns the time at the stop marker (X2

 

 

cursor).

Return Format

<value><NL>

 

 

<value> ::= time at the stop marker in NR3 format

See Also

"Introduction to :MARKer Commands" on page 204

 

 

"Introduction to :MEASure Commands" on page 218

 

 

":MARKer:X1Position" on page 206

 

 

":MARKer:X2Position" on page 208

 

 

":MARKer:XDELta" on page 210

 

 

":MEASure:TDELta" on page 480

 

 

":MEASure:TSTArt" on page 484

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

485

Page 485
Image 485