Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 373

:TRIGger:SPI:CLOCk:SLOPe

(see page 530)
Command Syntax :TRIGger:SPI:CLOCk:SLOPe <slope>
<slope> ::= {NEGative | POSitive}
The :TRIGger:SPI:CLOCk:SLOPe command specifies the rising edge
(POSitive) or falling edge (NEGative) of the SPI clock source that will
clock in the data.
Query Syntax :TRIGger:SPI:CLOCk:SLOPe?
The :TRIGger:SPI:CLOCk:SLOPe? query returns the current SPI clock
source slope.
Return Format <slope><NL>
<slope> ::= {NEG | POS}
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:SPI:CLOCk:TIMeout" on page 374
":TRIGger:SPI:SOURce:CLOCk" on page 378