Commands by Subsystem 3

:TRIGger:SPI:CLOCk:SLOPe

(see page 530)

Command Syntax :TRIGger:SPI:CLOCk:SLOPe <slope>

<slope> ::= {NEGative POSitive}

The :TRIGger:SPI:CLOCk:SLOPe command specifies the rising edge (POSitive) or falling edge (NEGative) of the SPI clock source that will clock in the data.

Query Syntax :TRIGger:SPI:CLOCk:SLOPe?

The :TRIGger:SPI:CLOCk:SLOPe? query returns the current SPI clock source slope.

Return Format <slope><NL>

<slope> ::= {NEG POS}

See Also "Introduction to :TRIGger Commands" on page 313

":TRIGger:SPI:CLOCk:TIMeout" on page 374

":TRIGger:SPI:SOURce:CLOCk" on page 378