3Commands by Subsystem

Table 50 :MEASure Commands Summary (continued)

Command

Query

Options and Query Returns

 

 

 

n/a

:MEASure:VTIMe?

<vtime> ::= displayed time from

 

<vtime>[,<source>]

trigger in seconds in NR3 format

 

(see page 253)

<return_value> ::= voltage at the

 

 

specified time in NR3 format

 

 

<source>

::= {CHANnel<n>

 

 

FUNCtion

MATH}

 

 

<n> ::= 1-2 or 1-4 in NR1 format

 

 

 

 

:MEASure:VTOP

:MEASure:VTOP?

<source>

::= {CHANnel<n>

[<source>] (see

[<source>] (see

FUNCtion

MATH}

page 254)

page 254)

<n> ::= 1-2 or 1-4 in NR1 format

 

 

<return_value> ::= voltage at the

 

 

top of the waveform in NR3 format

 

 

 

 

:MEASure:XMAX

:MEASure:XMAX?

<source>

::= {CHANnel<n>

[<source>] (see

[<source>] (see

FUNCtion

MATH}

page 255)

page 255)

<n> ::= 1-2 or 1-4 in NR1 format

 

 

<return_value> ::= horizontal

 

 

value of

the maximum in NR3

 

 

format

 

 

 

 

 

:MEASure:XMIN

:MEASure:XMIN?

<source>

::= {CHANnel<n>

[<source>] (see

[<source>] (see

FUNCtion

MATH}

page 256)

page 256)

<n> ::= 1-2 or 1-4 in NR1 format

 

 

<return_value> ::= horizontal

 

 

value of

the maximum in NR3

 

 

format

 

 

 

 

 

Introduction to The commands in the MEASure subsystem are used to make parametric :MEASure measurements on displayed waveforms.

Commands

Measurement Setup

To make a measurement, the portion of the waveform required for that measurement must be displayed on the oscilloscope screen.

Measurement Type

Portion of waveform that must be displayed

 

 

period, duty cycle, or frequency

at least one complete cycle

 

 

pulse width

the entire pulse

 

 

rise time

rising edge, top and bottom of pulse

 

 

fall time

falling edge, top and bottom of pulse

 

 

Measurement Error

218

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 218
Image 218
Agilent Technologies 5000 Series manual Measurement Setup, Measurement Error