142 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

:CHANnel<n> Commands

Control all oscilloscope functions associated with individual analog

channels or groups of channels. See "Introduction to :CHANnel<n>

Commands" on page 143.

Table 44 :CHANnel<n> Commands Summary

Command Query Options and Query Returns
:CHANnel<n>:BWLimit
{{0 | OFF} | {1 |
ON}} (see page 145)
:CHANnel<n>:BWLimit?
(see page 145)
{0 | 1}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:COUPling
<coupling> (see
page 146)
:CHANnel<n>:COUPling?
(see page 146)
<coupling> ::= {AC | DC}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:DISPlay
{{0 | OFF} | {1 |
ON}} (see page 147)
:CHANnel<n>:DISPlay?
(see page 147)
{0 | 1}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:IMPedance
<impedance> (see
page 148)
:CHANnel<n>:IMPedance
? (see page 148)
<impedance> ::= {ONEMeg | FIFTy}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:INVert
{{0 | OFF} | {1 |
ON}} (see page 149)
:CHANnel<n>:INVert?
(see page 149)
{0 | 1}
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:LABel
<string> (see
page 150)
:CHANnel<n>:LABel?
(see page 150)
<string> ::= any series of 6 or
less ASCII characters enclosed in
quotation marks
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:OFFSet
<offset>[suffix] (see
page 151)
:CHANnel<n>:OFFSet?
(see page 151)
<offset> ::= Vertical offset
value in NR3 format
[suffix] ::= {V | mV}
<n> ::= 1-2 or 1-4; in NR1 format
:CHANnel<n>:PROBe
<attenuation> (see
page 152)
:CHANnel<n>:PROBe?
(see page 152)
<attenuation> ::= Probe
attenuation ratio in NR3 format
<n> ::= 1-2 or 1-4r in NR1 format
n/a :CHANnel<n>:PROBe:ID?
(see page 153)
<probe id> ::= unquoted ASCII
string up to 11 characters
<n> ::= 1-2 or 1-4 in NR1 format
:CHANnel<n>:PROBe:SKE
W <skew_value> (see
page 154)
:CHANnel<n>:PROBe:SKE
W? (see page 154)
<skew_value> ::= -100 ns to +100
ns in NR3 format
<n> ::= 1-2 or 1-4 in NR1 format