Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 93

:AUToscale:CHANnels

(see page 530)
Command Syntax :AUToscale:CHANnels <value>
<value> ::= {ALL | DISPlayed}
The :AUTOscale:CHANnels command specifies which channels will be
displayed on subsequent :AUToscales.
When ALL is selected, all channels that meet the requirements of
:AUToscale will be displayed.
When DISPlayed is selected, only the channels that are turned on are
autoscaled.
Use the :VIEW or :BLANk root commands to turn channels on or off.
Query Syntax :AUToscale:CHANnels?
The :AUToscale:CHANnels? query returns the autoscale channels setting.
Return Format <value><NL>
<value> ::= {ALL | DISP}
See Also "Introduction to Root (:) Commands" on page 88
":AUToscale" on page 90
":AUToscale:AMODE" on page 92
":VIEW" on page 122
":BLANk" on page 94