66 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

*ESE (Standard Event Status Enable)

(see page 530)
Command Syntax *ESE <mask_argument>
<mask_argument> ::= integer from 0 to 255
The *ESE common command sets the bits in the Standard Event Status
Enable Register. The Standard Event Status Enable Register contains a
mask value for the bits to be enabled in the Standard Event Status
Register. A "1" in the Standard Event Status Enable Register enables the
corresponding bit in the Standard Event Status Register. A zero disables
the bit.
Table 29 Standard Event Status Enable (ESE)
Bit Name Description When Set (1 = High = True), Enables:
7 PON Power On Event when an OFF to ON transition occurs.
6 URQ User Request Event when a front-panel key is pressed.
5 CME Command Error Event when a command error is detected.
4 EXE Execution Error Event when an execution error is detected.
3 DDE Device Dependent Error Event when a device-dependent error is detected.
2 QYE Query Error Event when a query error is detected.
7R(6%ELWLQ
6WDWXV%\WH
5HJLVWHU
6WDQGDUG(YHQW
6WDWXV(QDEOH
0DVN5HJLVWHU
6WDQGDUG(YHQW
6WDWXV5HJLVWHU
(65"
(6(
(6("
321 854 (;( ''( 4<( 54/ 23&&0(

25
