Commands by Subsystem 3

:TRIGger:DURation:LESSthan

(see page 530)

Command Syntax :TRIGger:DURation:LESSthan <less than time>[<suffix>]

<less than time> ::= maximum trigger duration in seconds (5 ns - 10 seconds) in NR3 format

<suffix> ::= {s ms us ns ps}

The :TRIGger:DURation:LESSthan command sets the maximum duration for the defined pattern when :TRIGger:DURation:QUALifier is set to LESSthan.

Query Syntax :TRIGger:DURation:LESSthan?

The :TRIGger:DURation:LESSthan? query returns the duration time for the defined pattern.

Return Format <less than time><NL>

See Also "Introduction to :TRIGger Commands" on page 313

":TRIGger:DURation:PATTern" on page 338

":TRIGger:DURation:QUALifier" on page 339

":TRIGger:MODE" on page 319

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

337