Commands by Subsystem 3

:TRIGger:SWEep

Command Syntax

N O T E

(see page 530)

:TRIGger:SWEep <sweep>

<sweep> ::= {AUTO NORMal}

The :TRIGger:SWEep command selects the trigger sweep mode.

When AUTO sweep mode is selected, a baseline is displayed in the absence of a signal. If a signal is present but the oscilloscope is not triggered, the unsynchronized signal is displayed instead of a baseline.

When NORMal sweep mode is selected and no trigger is present, the instrument does not sweep, and the data acquired on the previous trigger remains on the screen.

This feature is called "Mode" on the instrument's front panel.

Query Syntax

:TRIGger:SWEep?

 

The :TRIGger:SWEep? query returns the current trigger sweep mode.

Return Format

<sweep><NL>

 

<sweep> ::= current trigger sweep mode

See Also

"Introduction to :TRIGger Commands" on page 313