5Obsolete and Discontinued Commands

:MEASure:UPPer

 

 

(see page 530)

Command Syntax

:MEASure:UPPer <value>

 

 

The :MEASure:UPPer command sets the upper measurement threshold

 

 

value. This value and the LOWer value represent absolute values when the

 

 

thresholds are ABSolute and percentage when the thresholds are PERCent

 

 

as defined by the :MEASure:DEFine THResholds command.

 

 

The :MEASure:UPPer command is obsolete and is provided for backward compatibility to

N O T E

 

 

previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 222)

 

 

 

 

instead.

 

 

Query Syntax

:MEASure:UPPer?

 

 

The :MEASure:UPPer? query returns the current upper threshold level.

Return Format

<value><NL>

 

 

<value> ::= the user-defined upper threshold in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

 

":MEASure:LOWer" on page 478

 

 

":MEASure:THResholds" on page 481

488

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 488
Image 488
Agilent Technologies 5000 Series manual MEASureUPPer, Instead