Commands by Subsystem 3

:ACQuire Commands

Set the parameters for acquiring and storing data. See "Introduction to :ACQuire Commands" on page 123.

Table 42 :ACQuire Commands Summary

Command

Query

Options and Query Returns

 

 

 

n/a

:ACQuire:AALias? (see

{1 0}

 

page 125)

 

 

 

 

:ACQuire:COMPlete

:ACQuire:COMPlete?

<complete> ::= 100; an integer in

<complete> (see

(see page 126)

NR1 format

page 126)

 

 

 

 

 

:ACQuire:COUNt

:ACQuire:COUNt? (see

<count> ::= an integer from 1 to

<count> (see

page 127)

65536 in NR1 format

page 127)

 

 

 

 

 

:ACQuire:DAALias

:ACQuire:DAALias?

<mode> ::= {DISable AUTO}

<mode> (see page 128)

(see page 128)

 

 

 

 

:ACQuire:MODE <mode>

:ACQuire:MODE? (see

<mode> ::= {RTIMe ETIMe}

(see page 129)

page 129)

 

 

 

 

n/a

:ACQuire:POINts? (see

<# points> ::= an integer in NR1

 

page 130)

format

 

 

 

n/a

:ACQuire:SRATe? (see

<sample_rate> ::= sample rate

 

page 131)

(samples/s) in NR3 format

 

 

 

:ACQuire:TYPE <type>

:ACQuire:TYPE? (see

<type> ::= {NORMal AVERage

(see page 132)

page 132)

HRESolution PEAK}

 

 

 

Introduction to The ACQuire subsystem controls the way in which waveforms are :ACQuire acquired. These acquisition types are available: normal, averaging, peak

Commands detect, and high resolution. Two acquisition modes are available: real- time mode, and equivalent- time mode.

Normal

The :ACQuire:TYPE NORMal command sets the oscilloscope in the normal acquisition mode. For the majority of user models and signals, NORMal mode yields the best oscilloscope picture of the waveform.

Averaging

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

123