3Commands by Subsystem

:ACQuire:DAALias

 

(see page 530)

Command Syntax

:ACQuire:DAALias <mode>

 

<mode> ::= {DISable AUTO}

 

The :ACQuire:DAALias command sets the disable anti- alias mode of the

 

oscilloscope.

 

When set to DISable, anti- alias is always disabled. This is good for cases

 

where dithered data is not desired.

 

When set to AUTO, the oscilloscope turns off anti- alias control as needed.

 

Such cases are when the FFT or differentiate math functions are silent.

 

The :DIGitize command always turns off the anti- alias control as well.

Query Syntax

:ACQuire:DAALias?

 

The :ACQuire:DAALias? query returns the oscilloscope's current disable

 

anti- alias mode setting.

Return Format

<mode><NL>

 

<mode> ::= {DIS AUTO}

See Also

"Introduction to :ACQuire Commands" on page 123

 

":ACQuire:AALias" on page 125