336 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

:TRIGger:DURation:GREaterthan

(see page 530)
Command Syntax :TRIGger:DURation:GREaterthan <greater than time>[<suffix>]
<greater than time> ::= minimum trigger duration in seconds
(5 ns - 10 seconds) in NR3 format
<suffix> ::= {s | ms | us | ns | ps }
The :TRIGger:DURation:GREaterthan command sets the minimum duration
for the defined pattern when :TRIGger:DURation:QUALifier is set to
GREaterthan. The command also sets the timeout value when the
:TRIGger:DURation:QUALifier is set to TIMeout.
Query Syntax :TRIGger:DURation:GREaterthan?
The :TRIGger:DURation:GREaterthan? query returns the minimum duration
time for the defined pattern.
Return Format <greater than time><NL>
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:DURation:PATTern" on page 338
":TRIGger:DURation:QUALifier" on page 339
":TRIGger:MODE" on page 319