3Commands by Subsystem

:TRIGger:IIC:PATTern:ADDRess

 

(see page 530)

Command Syntax

:TRIGger:IIC:PATTern:ADDRess <value>

 

<value> ::= integer or <string>

 

<string> ::= "0xnn" where n ::= {0,..,9 A,..,F}

 

The :TRIGger:IIC:PATTern:ADDRess command sets the address for IIC

 

data.The address can range from 0x00 to 0x7F (7- bit) or 0x3FF (10- bit)

 

hexadecimal. Use the don't care address (- 1 or 0xFFFFFFFF) to ignore the

 

address value.

Query Syntax

:TRIGger:IIC:PATTern:ADDRess?

 

The :TRIGger:IIC:PATTern:ADDRess? query returns the current address for

 

IIC data.

Return Format

<value><NL>

 

<value> ::= integer

See Also

"Introduction to :TRIGger Commands" on page 313

 

":TRIGger:IIC:PATTern:DATA" on page 357

 

":TRIGger:IIC:PATTern:DATa2" on page 358

 

":TRIGger:IIC:TRIGger[:TYPE]" on page 362

356

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 356
Image 356
Agilent Technologies 5000 Series manual TRIGgerIICPATTernADDRess, TRIGgerIICPATTernDATA on, TRIGgerIICPATTernDATa2 on