3Commands by Subsystem

:MEASure:VRMS

(see page 530)

Command Syntax :MEASure:VRMS [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :MEASure:VRMS command installs a screen measurement and starts a dc RMS value measurement. If the optional source parameter is specified, the current source is modified.

N O T E

This command is not available if the source is FFT (Fast Fourier Transform).

 

Query Syntax

:MEASure:VRMS? [<source>]

 

The :MEASure:VRMS? query measures and outputs the dc RMS value of

 

the selected waveform. The dc RMS value is measured on an integral

 

number of periods of the displayed signal. If at least three edges are not

 

present, the oscilloscope computes the RMS value on all displayed data

 

points.

Return Format

<value><NL>

 

<value> ::= calculated dc RMS value in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

":MEASure:SOURce" on page 240

252

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 252
Image 252
Agilent Technologies 5000 Series manual MEASureVRMS, See Also