3Commands by Subsystem

CAN Id specification is set by the :TRIGger:CAN:PATTern:ID and:TRIGger:CAN:PATTern:ID:MODE commands.

CAN Data specification is set by the :TRIGger:CAN:PATTern:DATA command.

CAN Data Length Code is set by the :TRIGger:CAN:PATTern:DATA:LENGth command.

N O T E

SOF is the only valid selection for analog oscilloscopes. If the automotive CAN and LIN

serial decode option (Option AMS) has not been licensed, SOF is the only valid selection.

 

Query Syntax :TRIGger:CAN:TRIGger?

The :TRIGger:CAN:TRIGger? query returns the current CAN trigger on condition.

Return Format <condition><NL>

<condition> ::= {SOF DATA ERR IDD IDE IDR ALL OVER ACK}

Errors "- 241, Hardware missing" on page 501

See Also "Introduction to :TRIGger Commands" on page 313

":TRIGger:MODE" on page 319

":TRIGger:CAN:PATTern:DATA" on page 326

":TRIGger:CAN:PATTern:DATA:LENGth" on page 327

":TRIGger:CAN:PATTern:ID" on page 328

":TRIGger:CAN:PATTern:ID:MODE" on page 329

":TRIGger:CAN:SIGNal:DEFinition" on page 496

":TRIGger:CAN:SOURce" on page 332

334

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 334
Image 334
Agilent Technologies 5000 Series manual Query Syntax TRIGgerCANTRIGger?