3Commands by Subsystem

:AUToscale

(see page 530)

Command Syntax :AUToscale

:AUToscale [<source>[,..,<source>]] <source> ::= CHANnel<n>

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The <source> parameter may be repeated up to 5 times.

The :AUToscale command evaluates all input signals and sets the correct conditions to display the signals. This is the same as pressing the Autoscale key on the front panel.

If one or more sources are specified, those specified sources will be enabled and all others blanked. The autoscale channels mode (see ":AUToscale:CHANnels" on page 93) is set to DISPlayed channels. Then, the autoscale is performed.

When the :AUToscale command is sent, the following conditions are affected and actions are taken:

Thresholds.

Channels with activity around the trigger point are turned on, others are turned off.

Channels are reordered on screen; analog channel 1 first, followed by the remaining analog channels.

Delay is set to 0 seconds.

Time/Div.

The :AUToscale command does not affect the following conditions:

Label names.

Trigger conditioning.

The :AUToscale command turns off the following items:

Cursors.

Measurements.

Trace memories.

Delayed time base mode.

For further information on :AUToscale, see the User's Guide.

See Also "Introduction to Root (:) Commands" on page 88

":AUToscale:CHANnels" on page 93

90

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference