340 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
3Commands by Subsystem

:TRIGger:DURation:RANGe

(see page 530)
Command Syntax :TRIGger:DURation:RANGe <greater than time>[<suffix>],
<less than time>[<suffix>]
<greater than time> ::= minimum duration in seconds
(10 ns - 9.99 seconds) in NR3 format
<less than time> ::= maximum duration in seconds
(15 ns - 10 seconds) in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:DURation:RANGe command sets the duration for the defined
pattern when the :TRIGger:DURation:QUALifier command is set to INRange
or OUTRange.
Query Syntax :TRIGger:DURation:RANGe?
The :TRIGger:DURation:RANGe? query returns the duration time for the
defined pattern.
Return Format <greater than time>,<less than time><NL>
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:DURation:PATTern" on page 338
":TRIGger:DURation:QUALifier" on page 339
":TRIGger:MODE" on page 319
NOTE If you set the minimum duration longer than the maximum duration, the order of the
parameters is automatically reversed.