Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 237

:MEASure:RISetime

(see page 530)
Command Syntax :MEASure: RISetime [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:RISetime command installs a screen measurement and
starts a rise- time measurement. If the optional source parameter is
specified, the current source is modified.
Query Syntax :MEASure: RISetime? [<source>]
The :MEASure:RISetime? query measures and outputs the rise time of the
displayed rising (positive- going) edge closest to the trigger reference. For
maximum measurement accuracy, set the sweep speed as fast as possible
while leaving the leading edge of the waveform on the display. The rise
time is determined by measuring the time at the lower threshold of the
rising edge and the time at the upper threshold of the rising edge, then
calculating the rise time with the following formula:
rise time = time at upper threshold - time at lower threshold
Return Format <value><NL>
<value> ::= rise time in seconds in NR3 format
See Also "Introduction to :MEASure Commands" on page 218
":MEASure:SOURce" on page 240
":MEASure:FALLtime" on page 228
NOTE This command is not available if the source is FFT (Fast Fourier Transform).