Commands by Subsystem 3

:MEASure:RISetime

(see page 530)

Command Syntax :MEASure: RISetime [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :MEASure:RISetime command installs a screen measurement and starts a rise- time measurement. If the optional source parameter is specified, the current source is modified.

N O T E

This command is not available if the source is FFT (Fast Fourier Transform).

 

Query Syntax

:MEASure: RISetime? [<source>]

 

The :MEASure:RISetime? query measures and outputs the rise time of the

 

displayed rising (positive- going) edge closest to the trigger reference. For

 

maximum measurement accuracy, set the sweep speed as fast as possible

 

while leaving the leading edge of the waveform on the display. The rise

 

time is determined by measuring the time at the lower threshold of the

 

rising edge and the time at the upper threshold of the rising edge, then

 

calculating the rise time with the following formula:

 

rise time = time at upper threshold - time at lower threshold

Return Format

<value><NL>

 

<value> ::= rise time in seconds in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

":MEASure:SOURce" on page 240

 

":MEASure:FALLtime" on page 228

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

237