522 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
7Status Reporting

Operation Status Condition Register (:OPERegister:CONDition)

This register hosts the RUN bit (bit 3), the WAIT TRIG bit (bit 5), the
OVLR bit (bit 11), and the HWE bit (bit 12).
The :OPERegister:CONDition? query returns the value of the Operation
Status Condition Register.
The HWE bit (bit 12) comes from the Hardware Event Registers.
The RUN bit is set whenever the instrument is not stopped.
The WAIT TRIG bit is set by the Trigger Armed Event Register and
indicates that the trigger is armed.
The OVLR bit is set whenever a 50 input overload occurs.