3Commands by Subsystem

:MEASure:VMIN

(see page 530)

Command Syntax :MEASure:VMIN [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :MEASure:VMIN command installs a screen measurement and starts a minimum vertical value measurement. If the optional source parameter is specified, the current source is modified.

Query Syntax :MEASure:VMIN? [<source>]

The :MEASure:VMIN? query measures and outputs the minimum vertical value present on the selected waveform.

Return Format <value><NL>

<value> ::= minimum vertical value of the selected waveform in NR3 format

See Also "Introduction to :MEASure Commands" on page 218

":MEASure:SOURce" on page 240

":MEASure:VBASe" on page 248

":MEASure:VMAX" on page 249

":MEASure:VPP" on page 251

250

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 250