5Obsolete and Discontinued Commands

:EXTernal:INPut

Command Syntax

N O T E

(see page 530)

:EXTernal:INPut <impedance>

<impedance> ::= {ONEMeg FIFTy}

The :EXTernal:IMPedance command selects the input impedance setting for the external trigger. The legal values for this command are ONEMeg (1 M) and FIFTy (50).

The :EXTernal:INPut command is an obsolete command provided for compatibility to previous oscilloscopes. Use the :EXTernal:IMPedance command (see page 174) instead.

Query Syntax :EXTernal:INPut?

The :EXTernal:INPut? query returns the current input impedance setting for the external trigger.

Return Format <impedance value><NL>

<impedance value> ::= {ONEM FIFT}

See Also "Introduction to :EXTernal Trigger Commands" on page 171

"Introduction to :TRIGger Commands" on page 313

":CHANnel<n>:IMPedance" on page 148

468

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 468
Image 468
Agilent Technologies 5000 Series manual EXTernalINPut