Commands by Subsystem 3

:TRIGger:UART:TYPE

(see page 530)

Command Syntax :TRIGger:UART:TYPE <value>

<value> ::= {RSTArt RSTop RDATa RD1 RD0 RDX PARityerror

TSTArt TSTOp TDATa TD1 TD0 TDX}

The :TRIGger:UART:TYPE command selects the UART trigger type.

When one of the RD or TD types is selected, the :TRIGger:UART:DATA and :TRIGger:UART:QUALifier commands are used to specify the data value and comparison operator.

The RD1, RD0, RDX, TD1, TD0, and TDX types (for triggering on data and alert bit values) are only valid when a 9- bit width has been selected.

Query Syntax :TRIGger:UART:TYPE?

The :TRIGger:UART:TYPE? query returns the current UART trigger data value.

Return Format <value><NL>

<value> ::= {RSTA RST RDAT RD1 RD0 RDX PAR TSTA TSTO TDAT TD1 TD0 TDX}

See Also "Introduction to :TRIGger Commands" on page 313

":TRIGger:MODE" on page 319

":TRIGger:UART:DATA" on page 392

":TRIGger:UART:QUALifier" on page 396

":TRIGger:UART:WIDTh" on page 400

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

399

Page 399
Image 399
Agilent Technologies 5000 Series manual TRIGgerUARTTYPE, Return Format valueNL