Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 347

:TRIGger:GLITch Commands

Table 60 :TRIGger:GLITch Commands Summary

Command Query Options and Query Returns
:TRIGger:GLITch:GREat
erthan <greater than
time>[suffix] (see
page 348)
:TRIGger:GLITch:GREat
erthan? (see
page 348)
<greater than time> ::=
floating-point number from 5 ns
to 10 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:GLITch:LESSt
han <less than
time>[suffix] (see
page 349)
:TRIGger:GLITch:LESSt
han? (see page 349)
<less than time> ::=
floating-point number from 5 ns
to 10 seconds in NR3 format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:GLITch:LEVel
<level> [<source>]
(see page 350)
:TRIGger:GLITch:LEVel
? (see page 350)
For internal triggers, <level>
::= .75 x full-scale voltage from
center screen in NR3 format.
For external triggers, <level>
::= 2 volts with probe
attenuation at 1:1 in NR3 format.
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format
:TRIGger:GLITch:POLar
ity <polarity> (see
page 351)
:TRIGger:GLITch:POLar
ity? (see page 351)
<polarity> ::= {POSitive |
NEGative}
:TRIGger:GLITch:QUALi
fier <qualifier> (see
page 352)
:TRIGger:GLITch:QUALi
fier? (see page 352)
<qualifier> ::= {GREaterthan |
LESSthan | RANGe}
:TRIGger:GLITch:RANGe
<greater than
time>[suffix], <less
than time>[suffix]
(see page 353)
:TRIGger:GLITch:RANGe
? (see page 353)
<greater than time> ::= start
time from 10 ns to 9.99 seconds
in NR3 format
<less than time> ::= stop time
from 15 ns to 10 seconds in NR3
format
[suffix] ::= {s | ms | us | ns |
ps}
:TRIGger:GLITch:SOURc
e <source> (see
page 354)
:TRIGger:GLITch:SOURc
e? (see page 354)
<source> ::= {CHANnel<n> |
EXTernal}
<n> ::= 1-2 or 1-4 in NR1 format