Commands by Subsystem 3

:EXTernal:PROBe

(see page 530)

Command Syntax :EXTernal:PROBe <attenuation>

<attenuation> ::= probe attenuation ratio in NR3 format

The :EXTernal:PROBe command specifies the probe attenuation factor for the external trigger. The probe attenuation factor may be 0.1 to 1000. This command does not change the actual input sensitivity of the oscilloscope. It changes the reference constants for scaling the display factors and for setting trigger levels.

If an AutoProbe probe is connected to the oscilloscope, the attenuation value cannot be changed from the sensed value. Attempting to set the oscilloscope to an attenuation value other than the sensed value produces an error.

Query Syntax :EXTernal:PROBe?

The :EXTernal:PROBe? query returns the current probe attenuation factor for the external trigger.

Return Format <attenuation><NL>

<attenuation> ::= probe attenuation ratio in NR3 format

See Also "Introduction to :EXTernal Trigger Commands" on page 171

":EXTernal:RANGe" on page 179

"Introduction to :TRIGger Commands" on page 313

":CHANnel<n>:PROBe" on page 152

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

175