Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 365

:TRIGger:LIN:ID

(see page 530)
Command Syntax :TRIGger:LIN:ID <value>
<value> ::= 7-bit integer in decimal, <nondecimal>, or <string>
from 0-63 or 0x00-0x3f
<nondecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<nondecimal> ::= #Bnn...n where n ::= {0 | 1} for binary
<string> ::= "0xnn" where n ::= {0,..,9 | A,..,F} for hexadecimal
The :TRIGger:LIN:ID command defines the LIN identifier searched for in
each CAN message when the LIN trigger mode is set to frame ID.
Setting the ID to a value of "- 1" results in "0xXX" which is equivalent to
all IDs.
Query Syntax :TRIGger:LIN:ID?
The :TRIGger:LIN:ID? query returns the current LIN identifier setting.
Return Format <value><NL>
<value> ::= integer in decimal
Errors "- 241, Hardware missing" on page 501
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:MODE" on page 319
":TRIGger:LIN:TRIGger" on page 371
":TRIGger:LIN:SIGNal:DEFinition" on page 497
":TRIGger:LIN:SOURce" on page 368
NOTE This command is only valid when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.