Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 329

:TRIGger:CAN:PATTern:ID:MODE

(see page 530)
Command Syntax :TRIGger:CAN:PATTern:ID:MODE <value>
<value> ::= {STANdard | EXTended}
The :TRIGger:CAN:PATTern:ID:MODE command sets the CAN identifier
mode. STANdard selects the standard 11-bit identifier. EXTended selects
the extended 29-bit identifier. The CAN identifier is set by the
:TRIGger:CAN:PATTern:ID command.
Query Syntax :TRIGger:CAN:PATTern:ID:MODE?
The :TRIGger:CAN:PATTern:ID:MODE? query returns the current setting of
the CAN identifier mode.
Return Format <value><NL>
<value> ::= {STAN | EXT}
Errors "- 241, Hardware missing" on page 501
See Also "Introduction to :TRIGger Commands" on page 313
":TRIGger:MODE" on page 319
":TRIGger:CAN:PATTern:DATA" on page 326
":TRIGger:CAN:PATTern:DATA:LENGth" on page 327
":TRIGger:CAN:PATTern:ID" on page 328
NOTE This command is only valid when the automotive CAN and LIN serial decode option (Option
AMS) has been licensed.