3Commands by Subsystem

:MEASure:PHASe

(see page 530)

Command Syntax :MEASure:PHASe [<source1>][,<source2>]

<source1>, <source2> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

 

The :MEASure:PHASe command places the instrument in the continuous

 

measurement mode and starts a phase measurement.

Query Syntax

:MEASure:PHASe? [<source1>][,<source2>]

 

The :MEASure:PHASe? query measures and returns the phase between the

 

specified sources.

 

A phase measurement is a combination of the period and delay

 

measurements. First, the period is measured on source1. Then the delay is

 

measured between source1 and source2. The edges used for delay are the

 

source1 rising edge used for the period measurement closest to the

 

horizontal reference and the rising edge on source 2. See :MEASure:DELay

 

for more detail on selecting the 2nd edge.

 

The phase is calculated as follows:

 

phase = (delay / period of input 1) x 360

Return Format

<value><NL>

 

<value> ::= the phase angle value in degrees in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

":MEASure:DELay" on page 225

 

":MEASure:PERiod" on page 233

 

":MEASure:SOURce" on page 240

234

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 234