3Commands by Subsystem

:MEASure:NWIDth

(see page 530)

Command Syntax :MEASure:NWIDth [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :MEASure:NWIDth command installs a screen measurement and starts a negative pulse width measurement. If the optional source parameter is specified, the current source is modified.

N O T E

This command is not available if the source is FFT (Fast Fourier Transform).

 

Query Syntax

:MEASure:NWIDth? [<source>]

 

The :MEASure:NWIDth? query measures and outputs the width of the

 

negative pulse on the screen closest to the trigger reference using the

 

midpoint between the upper and lower thresholds.

 

FOR the negative pulse closest to the trigger point:

 

width = (time at trailing rising edge - time at leading falling edge)

Return Format

<value><NL>

 

<value> ::= negative pulse width in seconds in NR3 format

See Also

"Introduction to :MEASure Commands" on page 218

 

":MEASure:SOURce" on page 240

 

":MEASure:PWIDth" on page 236

 

":MEASure:PERiod" on page 233