5Obsolete and Discontinued Commands

:TRIGger:TV:TVMode

(see page 530)
Command Syntax :TRIGger:TV:TVMode <mode>
<mode> ::= {FIEld1 | FIEld2 | AFIelds | ALINes | LINE | VERTical
| LFIeld1 | LFIeld2 | LALTernate | LVERtical}
The :TRIGger:TV:MODE command selects the TV trigger mode and field.
The LVERtical parameter is only available when :TRIGger:TV:STANdard is
GENeric. The LALTernate parameter is not available when
:TRIGger:TV:STANdard is GENeric (see page 386).
Old forms for <mode> are accepted:
Query Syntax :TRIGger:TV:TVMode?
The :TRIGger:TV:TVMode? query returns the TV trigger mode.
Return Format <value><NL>
<value> ::= {FIE1 | FIE2 | AFI | ALIN | LINE | VERT | LFI1 | LFI2
| LALT | LVER}
<mode> Old Forms Accepted
FIEld1 F1
FIEld2 F2
AFIeld ALLFields, ALLFLDS
ALINes ALLLines
LFIeld1 LINEF1, LINEFIELD1
LFIeld2 LINEF2, LINEFIELD2
LALTernate LINEAlt
LVERtical LINEVert
NOTE The :TRIGger:TV:TVMode command is an obsolete command provided for compatibility to