Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 253

:MEASure:VTIMe

(see page 530)
Query Syntax :MEASure:VTIMe? <vtime_argument>[,<source>]
<vtime_argument> ::= time from trigger in seconds
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n>::={1|2|3|4}forthefour channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VTIMe? query returns the value at a specified time on the
source specified with :MEASure:SOURce. The specified time must be on
the screen and is referenced to the trigger event. If the optional source
parameter is specified, the current source is modified.
Return Format <value><NL>
<value> ::= value at the specified time in NR3 format
See Also "Introduction to :MEASure Commands" on page 218
":MEASure:SOURce" on page 240
":MEASure:TEDGe" on page 242
":MEASure:TVALue" on page 244
NOTE This query is not available if the source is FFT (Fast Fourier Transform).