9Programming Examples

*communication with the computer interrupts data acquisition.

*Setting up the oscilloscope over the bus causes the data

*buffers to be cleared and internal hardware to be reconfigured.

*If a measurement is immediately requested there may not have

*been enough time for the data acquisition process to collect

*data and the results may not be accurate. An error value of

*9.9E+37 may be returned over the bus in this situation.

*/

iprintf(id, ":DIGITIZE CHAN1\n");

}

/*

*analyze

*------------------------------------------------------------------

*In this example we will do the following:

*- Save the system setup to a file for restoration at a later time.

*- Save the oscilloscope display to a file which can be printed.

*- Make single channel measurements.

*/

 

 

void analyze (void)

 

 

{

 

 

double frequency, vpp;

/* Measurements. */

double vdiv, off, sdiv, delay;

/* Calculated from preamble data. */

int i;

/* Loop counter. */

/* Array for setup string. */

 

 

unsigned char setup_string[SETUP_STR_SIZE];

int setup_size;

 

 

FILE *fp;

 

 

unsigned char image_data[IMG_SIZE];

/* Array for image data. */

int img_size;

 

 

/* SAVE_SYSTEM_SETUP - The :SYSTEM:SETUP? query returns a program

*message that contains the current state of the instrument. Its

*format is a definite-length binary block, for example,

*#800002204<setup string><NL>

*where the setup string is 2204 bytes in length.

*/

setup_size = SETUP_STR_SIZE;

/* Query and read setup string. */

ipromptf(id, ":SYSTEM:SETUP?\n", "%#b\n", &setup_size, setup_string); printf("Read setup string query (%d bytes).\n", setup_size);

/* Write setup string to file. */

fp = fopen ("c:\\scope\\config\\setup.dat", "wb");

setup_size = fwrite(setup_string, sizeof(unsigned char), setup_size, fp);

fclose (fp);

printf("Wrote setup string (%d bytes) to file.\n", setup_size);

/* RESTORE_SYSTEM_SETUP - Uploads a previously saved setup string

*to the oscilloscope.

*/

/* Read setup string from file. */

fp = fopen ("c:\\scope\\config\\setup.dat", "rb"); setup_size = fread (setup_string, sizeof(unsigned char),

SETUP_STR_SIZE, fp); fclose (fp);

554

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 554
Image 554