44 Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference
2Commands Quick Reference

Table 17 :TIMebase Commands Summary

Command Query Options and Query Returns
:TIMebase:MODE
<value> (see
page 304)
:TIMebase:MODE? (see
page 304)
<value> ::= {MAIN | WINDow | XY |
ROLL}
:TIMebase:POSition
<pos> (see page 305)
:TIMebase:POSition?
(see page 305)
<pos> ::= time from the trigger
event to the display reference
point in NR3 format
:TIMebase:RANGe
<range_value> (see
page 306)
:TIMebase:RANGe? (see
page 306)
<range_value> ::= 10 ns through
500 s in NR3 format
:TIMebase:REFerence
{LEFT | CENTer |
RIGHt} (see page 307)
:TIMebase:REFerence?
(see page 307)
<return_value> ::= {LEFT | CENTer
| RIGHt}
:TIMebase:SCALe
<scale_value> (see
page 308)
:TIMebase:SCALe? (see
page 308)
<scale_value> ::= scale value in
seconds in NR3 format
:TIMebase:VERNier {{0
| OFF} | {1 | ON}}
(see page 309)
:TIMebase:VERNier?
(see page 309)
{0 | 1}
:TIMebase:WINDow:POSi
tion <pos> (see
page 310)
:TIMebase:WINDow:POSi
tion? (see page 310)
<pos> ::= time from the trigger
event to the delayed view
reference point in NR3 format
:TIMebase:WINDow:RANG
e <range_value> (see
page 311)
:TIMebase:WINDow:RANG
e? (see page 311)
<range value> ::= range value in
seconds in NR3 format for the
delayed window
:TIMebase:WINDow:SCAL
e <scale_value> (see
page 312)
:TIMebase:WINDow:SCAL
e? (see page 312)
<scale_value> ::= scale value in
seconds in NR3 format for the
delayed window

Table 18 General :TRIGger Commands Summary

Command Query Options and Query Returns
:TRIGger:HFReject {{0
| OFF} | {1 | ON}}
(see page 317)
:TRIGger:HFReject?
(see page 317)
{0 | 1}
:TRIGger:HOLDoff
<holdoff_time> (see
page 318)
:TRIGger:HOLDoff?
(see page 318)
<holdoff_time> ::= 60 ns to 10 s
in NR3 format