3Commands by Subsystem

:MEASure:VAMPlitude

(see page 530)

Command Syntax :MEASure:VAMPlitude [<source>]

<source> ::= {CHANnel<n> FUNCtion MATH}

<n> ::= {1 2 3 4} for the four channel oscilloscope models <n> ::= {1 2} for the two channel oscilloscope models

The :MEASure:VAMPlitude command installs a screen measurement and starts a vertical amplitude measurement. If the optional source parameter is specified, the current source is modified.

Query Syntax :MEASure:VAMPlitude? [<source>]

The :MEASure:VAMPlitude? query measures and returns the vertical amplitude of the waveform. To determine the amplitude, the instrument measures Vtop and Vbase, then calculates the amplitude as follows:

vertical amplitude = Vtop - Vbase

Return Format <value><NL>

<value> ::= the amplitude of the selected waveform in NR3 format

See Also "Introduction to :MEASure Commands" on page 218

":MEASure:SOURce" on page 240

":MEASure:VBASe" on page 248

":MEASure:VTOP" on page 254

":MEASure:VPP" on page 251

246

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 246
Image 246
Agilent Technologies 5000 Series manual MEASureVAMPlitude