Obsolete and Discontinued Commands 5
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 481

:MEASure:THResholds

(see page 530)
Command Syntax :MEASure:THResholds {T1090 | T2080 | VOLTage}
The :MEASure:THResholds command selects the thresholds used when
making time measurements.
Query Syntax :MEASure:THResholds?
The :MEASure:THResholds? query returns the current thresholds selected
when making time measurements.
Return Format {T1090 | T2080 | VOLTage}<NL>
{T1090} uses the 10% and 90% levels of the selected waveform.
{T2080} uses the 20% and 80% levels of the selected waveform.
{VOLTage} uses the upper and lower voltage thresholds set by the
UPPer and LOWer commands on the selected waveform.
See Also "Introduction to :MEASure Commands" on page 218
":MEASure:LOWer" on page 478
":MEASure:UPPer" on page 488
NOTE The :MEASure:THResholds command is obsolete and is provided for backward
compatibility to previous oscilloscopes. Use the :MEASure:DEFine THResholds command
(see page 222) instead.