Obsolete and Discontinued Commands 5

:MEASure:THResholds

 

 

(see page 530)

Command Syntax

:MEASure:THResholds {T1090 T2080 VOLTage}

 

 

The :MEASure:THResholds command selects the thresholds used when

 

 

making time measurements.

 

 

The :MEASure:THResholds command is obsolete and is provided for backward

N O T E

 

 

compatibility to previous oscilloscopes. Use the :MEASure:DEFine THResholds command

 

 

 

 

(see page 222) instead.

 

 

Query Syntax

:MEASure:THResholds?

 

 

The :MEASure:THResholds? query returns the current thresholds selected

 

 

when making time measurements.

Return Format

{T1090 T2080 VOLTage}<NL>

 

 

{T1090} uses the 10% and 90% levels of the selected waveform.

 

 

{T2080} uses the 20% and 80% levels of the selected waveform.

 

 

{VOLTage} uses the upper and lower voltage thresholds set by the

 

 

UPPer and LOWer commands on the selected waveform.

See Also

"Introduction to :MEASure Commands" on page 218

 

 

":MEASure:LOWer" on page 478

 

 

":MEASure:UPPer" on page 488

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

481

Page 481
Image 481