Commands by Subsystem 3

:ACQuire:COUNt

(see page 530)

Command Syntax :ACQuire:COUNt <count>

<count> ::= integer in NR1 format

In averaging mode, the :ACQuire:COUNt command specifies the number of values to be averaged for each time bucket before the acquisition is considered to be complete for that time bucket. When :ACQuire:TYPE is set to AVERage, the count can be set to any value from 1 (smoothing) to 65536.

Query Syntax :ACQuire:COUNT?

The :ACQuire:COUNT? query returns the currently selected count value for averaging mode.

Return Format <count_argument><NL>

<count_argument> ::= an integer from 1 to 65536 in NR1 format

See Also "Introduction to :ACQuire Commands" on page 123

":ACQuire:TYPE" on page 132

":DIGitize" on page 96

":WAVeform:COUNt" on page 409