3Commands by Subsystem

:TRIGger:NREJect

 

(see page 530)

Command Syntax

:TRIGger:NREJect <value>

 

<value> ::= {{0 OFF} {1 ON}}

 

The :TRIGger:NREJect command turns the noise reject filter off and on.

 

When the noise reject filter is on, the trigger circuitry is less sensitive to

 

noise but may require a greater amplitude waveform to trigger the

 

oscilloscope. This command is not valid in TV trigger mode.

Query Syntax

:TRIGger:NREJect?

 

The :TRIGger:NREJect? query returns the current noise reject filter mode.

Return Format

<value><NL>

 

<value> ::= {0 1}

See Also

"Introduction to :TRIGger Commands" on page 313

320

Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference

Page 320
Image 320
Agilent Technologies 5000 Series manual TRIGgerNREJect, Query Syntax